TECHNICAL SPECIFICATIONS FOR PATIENT BED HEAD UNITThe Bedside Unit Medical Gas Installation and Electrical Installations must be made of extruded first quality aluminum profiles consisting of three closed sections where they can be pulled separately.
The main body of the Patient Bedside Unit should have a duct at the bottom of which 95x22 mm medical gas pipes will pass and the medical gas pipes coming from the unit flow must be connected to the sockets in the unit through the opening of 80x80 mm size for the passage of these pipes. brass blocks and medical gas sockets must be connected to each...